| Product name | 
			RN PET S5 | 
		
		
			| Temperature | 
			Technique: Thermistor probe (2.25K/10k); | 
		
		
			| Measuring and Alarm Range: 0.0℃~50℃; | 
		
		
			| Unit: Celsius (℃); | 
		
		
			| Resolution: 0.1℃ or 1F; | 
		
		
			| Accuracy: ±0.1℃ (exclusive probe); | 
		
		
			| Pulse Rate | 
			Measuring and Alarm Range: 25~250bpm; | 
		
		
			| Accuracy: ±3bpm; | 
		
		
			| Resolution: 1bpm; | 
		
		
			| ECG | 
			Lead mode: 3 Leads I,II,III; | 
		
		
			| Gain: 1/4 file (X0.25) 2.5mm/mV; 1/2 file (X0.50) 5.0mm/mV; 1 file (X1) 10mm/mV; 2 files (X2) 20mm/mV; | 
		
		
			| Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s; | 
		
		
			| Baseline Recovery: < 3s after defibrillation (Monitor mode), < 1s after defibrillation (Surgery mode); | 
		
		
			| Bandwidth: Diagnostic 0.05~100, Monitor 0.5~40Hz, Surgery 1~20Hz; | 
		
		
			| NIBP | 
			Method: Pulse wave oscillation method; | 
		
		
			| Measure mode: Manual, Auto, Continue; | 
		
		
			| Measure interval in auto mode: 1 file (X1) 10mm/mV; 2 files (X2) 20mm/mV; 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480min; | 
		
		
			| STAT mode cycle time: Keep 5 minutes; | 
		
		
			| 
			 Measure and Alarm Range: 
			Large Animal: SYS 30~280mmHg DIA 10~220mmHg MEAN 20~240mmHg, 
			Small Animal: SYS 30~220mmHg DIA 10~160mmHg MEAN 20~170mmHg, 
			Static pressure range: 0~300mmHg, 
			Static pressure accuracy: 3mmHg; 
			 | 
		
		
			| Maximum Mean error: 5mmHg; | 
		
		
			| Maximum Standard deviation: 8mmHg; | 
		
		
			| Static pressure accuracy: ±3mmHg; | 
		
		
			| Resolution: 1mmHg; | 
		
		
			| Accuracy: Maximum Mean error ± 5mmHg, Maximum Standard deviation<8mmHg; | 
		
		
			| Over pressure Protection: Large animal~300mmHg, Small animal~240mmHg; | 
		
		
			| Heart Rate | 
			Measure range: Large animal 15~300 bpm, Small animal 15~350 bpm; | 
		
		
			| Accuracy: 1% or 1bpm, whichever is greater; | 
		
		
			| Resolution: 1 bpm; | 
		
		
			| SpO2 | 
			Measurement Range: 0~100%; | 
		
		
			| Resolution: 1%; | 
		
		
			| Response Modes: Low, Medium, High; | 
		
		
			| Accuracy: 70%~100%: 2% <69%: unspecified; | 
		
		
			| EtCO2 | 
			Measure method: Non-dispersive infrared (NDIR); | 
		
		
			| Measure Range: 0~19.7% (0~150mmHg) 0~20kPa; | 
		
		
			| Resolution: 0.1 mmHg; | 
		
		
			| 
			 CO2 Accuracy: 0~40 mmHg, ±2 mmHg, 41~70 mmHg, ±5% of reading, 71~100 mmHg, ±8% of reading, 101~150 mmHg, 
			±10% of reading, at 760mmHg, ambient temperature of 25℃); 
			 | 
		
		
			| Respiratory Rate: Range:3-150BrPM, Accuracy:±1bpm; | 
		
		
			| Transportation and Storage | 
			Temperature: -20~50℃; | 
		
		
			| Humidity: 10%~90%(non-condensing); | 
		
		
			| Atmosphere pressure: 86KPa~106KPa; | 
		
		
			| Respiration | 
			Method:lmpedance between RA-LL; | 
		
		
			| Gain: x0.25, x0.50, x1, x2, x4; | 
		
		
			| Respiration Rate:0~120/0-150BrPM; | 
		
		
			| Sweep speed: 6.25 mm/s,12.5 mm/s,25mm/s; | 
		
		
			| Resolution:1 BrPM; | 
		
		
			| Apnea Alarm:10~40s; | 
		
		
			| Measure range: Large animal: 0-120 bpm; | 
		
		
			| Small animal:0-150 bpm; | 
		
		
			| Trend & Reviewing | 
			NlBP measurement reviewing: 1500groups; | 
		
		
			| Waveform review: 3 hours 600groups; | 
		
		
			| Alarm event: 1000 groups of parameter alarms events and associated parameter; |